| Wavelength Range |
|
|
| 190-740 nm, cutoff filter (2nd
order) 370 nm |
| Light Source |
|
Important
Note:
Version with detached flow cell with fiber optics is optionally
available (order no. A5141). |
| Deuterium; tungsten halogen
is optional |
| Bandwidth / Wavelength
Accuracy |
|
| Δ λ ≤ 8 nm /
± 2 nm |
| Noise / Drift (ASTM
E1657-94) |
|
| < 1x10-5 AU /
< 1.5x10-4 AU/h |
| Spectral Scan |
|
| stop-flow |
| Sensitivity |
|
| < 2 x 10-5 AU
at 240 nm, 1.0 s |
| Time Constants |
|
| 0.1/0.2/0.5/1.0/2.0/5.0/10.0
s |
| Analog Output
(Integrator) |
|
| ±1 V (scalable in 16
increments from 10-4 to 10 AU) |
| Analog Input
(wavelength control) |
|
| 10 mV/nm (0 to 10 V) |
| Miscellaneous |
|
| Control: RS-232, remote
control strip, touchpad; GLP Report: e.g. total lamp operation time,
lamp ignitions; Wavelength Verification: holmium oxide filter cell;
Display: LCD, 2 lines, 24 characters each; Universal Power Supply:
90-260 V, 47-63 Hz |
| Dimensions / Weight |
|
| 226 x 135 x 390 mm (W x H x
D) / 6.0 kg |
| Technical data is subject to change